Email format error
Email cannot be empty
Email already exists
6-20 characters(letters plus numbers only)
The password is inconsistent
Email format error
Email cannot be empty
Email does not exist
6-20 characters(letters plus numbers only)
The password is inconsistent
LTX05 500mA DR X ray machine
Product Configuration:
1.X-ray tube
2.Collimator
3. Radiographic table
4.High voltage generator
5.Micro console
6.High voltage cable
7.Double Grid
8.Vertical bucky stand
Technical parameters of main components:
|
Parameter |
LTX05 |
|
|
Intermittent mode |
100kV ,500mA,50kW |
|
|
Power Voltage |
Three-phase five wires 380V±38V |
|
|
tube current |
Radiography16mA~200mA |
Radiography16mA~630mA |
|
Output Voltage |
Radiography 40-125KVstep 1kV |
|
|
resistance power |
less than 0.5Ω |
|
|
Output power |
50kW |
|
|
Exposure Time |
Radiography 0.0025s~6.4s adjustable |
|
|
Nominal Frequency |
≥30kHz |
|
|
Anode speed |
2800r/min |
|
|
Anode heat capacity |
140KHU |
|
|
Focus |
Small focus:1.0mm Large focus:2.0mm |
|
|
L*W*H of bed surface |
2100mm×825mm×650mm |
|
|
Horizontal moving range |
≥250mm |
|
|
Vertical moving range |
≥825mm |
|
|
grid density |
≥40 lines/cm. |
|
|
grid ratio |
10:01 |
|
|
Focus Distance |
100cm |
|
|
Cassette size |
5″×7″ 8″×10″ 10″×12″ 11″×14″ 14″×17″ |
|
High Frequency High Voltage X-ray Generator
|
operating console |
|
|
Power |
AC380V ± 38V |
|
Radiography Tube Voltage |
40kV ~ 150kV |
|
Radiography Tube Current |
10mA ~ 630mA |
|
Time |
2ms ~ 6300ms |
|
Allowed Internal Resistance |
380V: 0.5Ω |
|
Allowed Frequency Variation |
50Hz/60Hz ± 1Hz |
|
Working Frequency |
≥30kHz |
X-ray Tube Assembly
|
Focus |
0.6mm/1.2mm |
|
Rotating Speed |
2800rpm |
|
Peek Tube Voltage |
150kV |
|
Small Focus |
20kW |
|
Large Focus |
50kW |
Collimator
|
Mode of adjustment |
Manual |
|
Power |
24VAC |
|
Inherent Filtration |
1.2mmAL |
Digital Detector Rayence 1717SGC/SCC
|
Scanning Area |
17 × 17 inches |
|
Spatial Resolution |
not less than 3.9lp/mm |
|
Total Pixel Matrix |
3,328(h) × 3,328(v) |
|
Pexel Pitch |
127 |
|
Image Preview Time |
not more than 2s |
DR Machine Frame
|
Range of horizontal movement of bed |
≥250mm |
|
Rang of Vertical movement of bed |
≥825mm |
|
Tabletop Size (Length × Width × Height) |
2100mm × 825mm × 655mm |
|
Mode of Operation |
Manual or electronic |
|
Load-bearing |
Less than 250kg |
Computer and LCD DCSM1F/ML2002
|
CPU |
≥2.8GHz |
|
Memory |
≥2GB |
|
Disk Capacity |
≥250GB |
|
Medical LCD |
19 inch |